Bio-Rad CHEF Mapper XA Analyzer System 1703670 F Power Module Tested

Score: 5.00 (votes: 1)
No reviews.
5.00
  • $495.00
Out of stock
Bio-Rad CHEF MAPPER XA Power Module
Bio-Rad CHEF PULSED FIELD ELECTROPHORESIS SYSTEM

GENUINE OEM Bio-Rad Equipment
Serial  801BR-2217 / System 1703670
Analyzer System F

TESTED Unit - See Description Below !
 
Other System Components Available As Well
Get A GOOD One -
at GetThatPart !

About This Item
Actual item from lot is shown. Only what is shown is included. TESTED Unit - pulled from working system and cleaned, inspected, and recertified for dependable daily use by our tech team. Unit was tested prior to parts harvest and determined to be working perfectly. Guaranteed functional for a period of 30 days.

This component is part of a larger system. We are offering this component individually for those who may need to replace only one component, but we do have the entire system as described below. See description below for specific testing, serial numbers, and other component specific information. Please ask us about the other compoents if you need them as well.

NOTE - The Full System Manual is attached for your convenience - see 'attachments'

Know What You are Buying

  • We typically take the time to post actual pictures of the unit you are buying
  • We also typically take the time to thoroughly test the units we are selling
  • Buy with CONFIDENCE from a reputable seller with 20 Years experience

DESCRIPTION

Component of the system being offered in this listing is  - Bio-Rad Chef Mapper XA / Power Module with Analyzer System F. System MPN is 1703670 and the serial on this component is 801BR-2217

In testing, this unit powers to paused status, all screens respond from there as anticipated. All components power and run as expected. The current parameters setup is shown but it can be programmed as you need.

System Description (for which this component is intended)

Bio-Rad CHEF MAPPER XA PULSED FIELD ELECTRPHORESIS SYSTEM

The CHEF Mapper system is based on two leading technologies, CHEF (clamped homogeneous electric fields)31 and PACE (programmed autonomously controlled electrodes).32 The system provides highly uniform, or homogeneous, electric fields within the gel, using an array of 24 electrodes, some of which are clamped, or held to intermediate potentials to eliminate lane distortion. Thus, lanes are straight. The system maintains uniform fields using patented Dynamic Regulation (US patent number 4,878,008). The electrodes sense changes in buffer conductivity due to buffer breakdown, buffer type, gel thickness, pH fluctuations, and temperature, and potentials are readjusted immediately to maintain uniform fields, thus insuring high resolution. In PACE, each electrode s voltage is controlled independently by firmware. Whereas other CHEF systems are limited to two vectors and a 120° pulse angle, the CHEF Mapper system allows up to 15 vectors per block with a total of 8 blocks, each vector with its own voltage, angle, and duration. Thus, the CHEF Mapper system may simulate virtually any pulsed field technique using homogeneous fields, including FIGE, CHEF, and RFGE. Advanced programmers may simulate OFAGE and other non-homogeneous field methods using a computer.

The CHEF Mapper system offers innovations beyond original PACE. For example, nonlinear switch time ramping allows linear separations for many sizes of DNA. Secondary pulses, or interrupts, unhinge DNAs from obstructions and permit faster separations. The CHEF Mapper system contains 5 years of protocols embedded on a microchip, eliminating trial and error in setting parameters. Given the size range you expect to separate, the embedded auto algorithm interrelates the sizes with 10 other variables, and provides the preferred operating conditions. Common gel and buffer conditions, and run temperature of 14 °C, are assumed. An Interactive Program Disc provides an extended version of the algorithm. This PC program allows you to vary gel, buffer, and temperature parameters and print out optimal conditions.

We offer a Bio-Rad Chef Mapper in good physical condition. The unit powers on fine. Compressor Kicks on and cools to temp.

The System will include the following components:

  • Bio-Rad CHEF Mapper Power Module
  • Bio-Rad Electrophoresis Cell with cable and lead
  • Bio-Rad Chiller Module
  • Bio-Rad Variable Speed Pump
  • Rom 2.1

This complete system sells for $22,000

 

CALL US with questions - Our business line is 410-374-1261

 

Thanks for looking!
    • SKU
      SKU104140
    • Weight
      50.0000 lbs
    • Dimensions (L x W x H)
      30 x 20 x 20 in
  • Manufacturer
    Bio-Rad
  • Product Added Date
    8/31/2019
  • Inventory Added Date
    8/31/2019
  • Condition Code
    Refurbished
  • Bin Location
    MH-RA1
  • P/N
    1703670
  • BB Condition Code
    Used
Customer reviews
Average rating:
5.00(Votes: 1)
Rating of votes (1)
5
 
1customers
4
 
 
0customers
3
 
 
0customers
2
 
 
0customers
1
 
 
0customers
 
There are no comments yet. Be the first to leave a comment
Please sign in to add review
Power Module 
Dimensions (W x D x H)55.9 x 34.5 x 30 cm
ConstructionAluminum chassis
Weight16 kg (35.3 lb)
Power supply350 V maximum, to allow maximum gradient of 9 V/cm, continuously adjustable; built in
Maximum current0.5 A
Allowable voltage gradients0.6–9 V/cm, in 0.1 V/cm increments
Battery back-upAll parameters in memory, up to 3 hr of interruption
Delayed startUp to 72 hr
Electrode potentialsDynamically regulated (feedback adjustment) +/- 0.5% F.S.
Program storage20 average size protocols
DisplayFluorescent, 2 lines x 40 characters per line, alphanumeric
Environmental 
Operating temperature50°F (10°C) to 90°F (32°C); 30–80% humidity
Storage temperature32°F (0°C) to 140°F (60°C); 10–90% humidity
Switching Functions 
Switching range50 ms to 18 hr
Switch angle variable0–360° (all electronic switching) in 0.5° increments
Multistate vector switchingUp to 15 vectors per pulse cycle, each definable by angle, voltage, and duration
Switch time rampsLinear, concave, or convex using hyperbolic function
Secondary pulsesDefined by voltage, frequency, angle, and duration
Field inversion (FIGE)Available with asymmetric forward, reverse voltages
Maximum program blocks8, with automatic execution
Maximum run time999 hr per block
Fuses3 A Slow-Blow; two each for AC line input
Electrophoresis Cell 
Dimensions (W x D x H)44.2 x 50.3 x 11.4 cm, horizontal format
ConstructionCover: vacuum-formed polycarbonate, base: injection-molded polycarbonate
LidSafety interlocked
Weight10.2 kg (22.5 lb)
Electrodes24, platinum (0.02" diameter)
Temperature monitoringVia precision temperature probe mounted in base of cell
Accessories Included 
Variable speed oscillating pump120 V, ground isolated; flow rate 1 liter/min, typical
Casting stand14 x 13 cm
Comb15-well comb and comb holder
Tygon tubing365 cm
Disposable sample plug mold50 slot
Yeast DNA standardS. cerevisiae YNN295
Chromosomal grade agarose5 g
Pulsed field Certified agarose5 g
Leveling bubble1
Cooling Module (optional) 
Weight14 kg (30.9 lb)
ConstructionAluminum
Dimensions (W x L x H)23 x 42 x 24 cm
Cooling capacity75 W of input power at 14°C
Operating range5–25°C
Total system weight41.7 kg (91.9 lb)

 

Footer navigation